Gas Chromatographs
http://www.sibel.info/en/gas-chromatographs.html#sigProGalleria89afd450ff
Application
Fast-acting portable automatic gas chromatographs GCS-02F and GCS-02FN are equipped with Electron Capture Detector (ECD) and Photo-Ionization Detector (PID) respectively, and are dezigned for qualitative and quantitative analysis within wide range of organic substances contained in gaseous and liquid samples, either in steady-state conditions or as a part of a mobile laboratory.
The main fields of application:
– detection and identification of explosives
– express drug analysis and identification, first of all, in field conditions
We’ve got the Type of Measuring Tool certificate RU.C.31.001.A № 26303 for chromatograph GCS-02FN.
Technical characteristics
Straight multicapillary columns of different types with stationary liquid phases (SE-30, SE-54, Carbowax 20M, OV-61) can be used with the chromatographs.
| Column efficiency: | from 1000 to 3000 theoretical plates (depending on the detector type) |
| Thermostats temperature range: | from 50 to 250°C |
| Warm-up period: | no more than 30 minutes at column temperature of 200°C |
| Time of continuous operation for GCS-02F: | 40 hours with fully charged (150 atm.) integrated gas cylinder (ECD) |
| Time of continuous operation for GCS-02FN: | unlimited |
| Power supply: | from main (100 ... 240V) or from external rechargeable battery (for exaple, car power supply) |
| Power consumption: : - warm up - operation |
200 W 100 W |
| Operating temperature: | +5 ... +40°C (relative humidity up to 80%) |
| Overall dimensions: | 462x344x184 mm |
| Weight in transportation case: | 14,5 kg |


Download detailed description in PDF.