Gas Chromatographs

Application

Fast-acting portable automatic gas chromatographs GCS-02F and GCS-02FN are equipped with Electron Capture Detector (ECD) and Photo-Ionization Detector (PID) respectively, and are dezigned for qualitative and quantitative analysis within wide range of organic substances contained in gaseous and liquid samples, either in steady-state conditions or as a part of a mobile laboratory.

The main fields of application:

– detection and identification of explosives
– express drug analysis and identification, first of all, in field conditions

We’ve got the Type of Measuring Tool certificate RU.C.31.001.A № 26303 for chromatograph GCS-02FN.

Technical characteristics

Straight multicapillary columns of different types with stationary liquid phases (SE-30, SE-54, Carbowax 20M, OV-61) can be used with the chromatographs.

Column efficiency: from 1000 to 3000 theoretical plates (depending on the detector type)
Thermostats temperature range: from 50 to 250°C
Warm-up period: no more than 30 minutes
at column temperature of 200°C
Time of continuous operation for GCS-02F: 40 hours with fully charged (150 atm.) integrated gas cylinder (ECD)
Time of continuous operation for GCS-02FN: unlimited
Power supply: from main (100 ... 240V) or from external rechargeable battery (for exaple, car power supply)
Power consumption: :

- warm up
- operation


200 W
100 W
Operating temperature: +5 ... +40°C (relative humidity up to 80%)
Overall dimensions: 462x344x184 mm
Weight in transportation case: 14,5 kg


Download detailed description in PDF.